Pacific Power Source LMX Series Operation Manual page 299

Programmable ac power source
Table of Contents

Advertisement

Command Syntax
Description
Parameters
Parameter Format
Example
Query Format
Returned Data Format
Query Example
Query Syntax
Description
Returned Data Format
Query Example
Query Syntax
Description
Returned Data Format
Query Example
Query Syntax
Description
Returned Data Format
Query Example
Command Syntax
Description
Parameters
Parameter Format
Example
Query Format
Returned Data Format
Query Example
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
LMX Series Power Source Operation Manual
[SOURce:]PULSe:DCYCle <nr1>, <opt>
This command sets the pulse duty cycle in percent. Minimum set
value allowed is 1, max is 99.
Duty Cycle %, Option: < NORmal | SATurate >
Default is NORmal if optional second parameter is omitted.
<nr1>, <cr>
PULS:DCYC 50
[SOURce:]PULSe:DCYCle?
<nr1>
PULS:DCYC?
50.0000
[SOURce:]PULSe:DCYCle:MINimum?
This command returns the minimum permissible set value for the
PULSE duty cycle.
<nr2>
PULS:DCYC:MIN?
0.0000
[SOURce:]PULSe:DCYCle:MAXimum?
This command returns the maximum permissible set valuefor the
PULSE duty cycle.
<nr2>
PULS:DCYC:MAX?
100.0000
[SOURce:]PULSe:DCYCle:DEFault?
This command returns the default set value for the PULSE duty cycle.
<nr2>
PULS:DCYC:DEF?
50.0000
[SOURce:]PULSe:PERiod <nr2>, <opt>
This command sets the pulse period in seconds.
Period, Option: < NORmal | SATurate >
Default is NORmal if optional second parameter is omitted.
<nr2>, <cr>
PULS:PER 12.8
[SOURce:]PULSe:PERiod?
<nr2>
PULS:PER?
12.8000
LMX SERIES™ OPERATION MANUAL
Remote Control Programming
SECTION 8:
Page 299 of 463

Advertisement

Table of Contents
loading

Table of Contents