Pacific Power Source LMX Series Operation Manual page 451

Programmable ac power source
Table of Contents

Advertisement

Error Source
Code
I/O Board
397
I/O Board
398
I/O Board
399
I/O Board
400
I/O Board
401
I/O Board
402
I/O Board
454
I/O Board
455
I/O Board
456
Controller
459
Controller
460
Controller
461
462
Controller
Controller
463
Controller
465
Controller
466
Controller
467
Controller
478
Controller
483
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
LMX Series Power Source Operation Manual
Category
Error Message
Internal
Invalid constants loaded
Error
from EEPROM.
Internal
Controller is not in
Error
operational node.
Error
Transformer coupling
circuits hardware fault.
Error
Transformer coupling
circuits not detected.
Error
Remote inhibit was issued
from remote interface.
Error
Forced fault.
Error
Controller is not in
operational mode.
Error
Missing controller node.
Error
Cannot read fault
information.
Error
Internal board fault.
Error
No waveform loaded.
Load waveform before
enabling arbitrary
waveform mode.
Error
Exceeded user RMS
current limit.
Error
Exceeded user active
power limit.
Error
Exceeded user apparent
power limit..
Internal
Internal McBSP
Error
communication CRC
error.
Internal
Internal sync error in
Error
CPU1.
Internal
Internal sync error in
Error
CPU2.
Error
Global fault input signal
detected.
Internal
Invalid node ID. Run unit
Error
discovery to recover.
LMX SERIES™ OPERATION MANUAL
SECTION 12: Warnings & Error Messages
Possible Cause
Configuration Error.
Coefficients found are
invalid.
An inconsistency was
detected in the power
source's internal
communication network.
An inconsistency was
detected in the power
source's internal
communication network.
An inconsistency was
detected in the power
source's internal
communication network.
An inconsistency was
detected in the power
source's internal
communication network.
Resolution
Sanitize unit to restore
factory defaults
Turn unit off, check
parallel cable connections
if present and then turn
unit back on. If problem
recurs, contact customer
service.
Turn unit off, check
parallel cable connections
if present and then turn
unit back on. If problem
recurs, contact customer
service.
Turn unit off, check
parallel cable connections
if present and then turn
unit back on. If problem
recurs, contact customer
service.
Turn unit off, check
parallel cable connections
if present and then turn
unit back on. If problem
recurs, contact customer
service.
Page 451 of 463