4. Enable the Output with no load connected
5. Press "Cal I. offset" soft key
6. Wait for the I offset coefficients for all phases to be calculated and displayed
Note: The Mode, Voltage and Current edit boxes do not affect this function.
11.4.3 Voltage Calibration - Gain
Figure 11-1: Voltage Calibration Equipment Setup – 1 or 3 Phase Mode – Phase A
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
LMX Series Power Source Operation Manual
DMM
LMX SERIES™ OPERATION MANUAL
Calibration
SECTION 11:
Page 432 of 463