8.4 Measurement Commands

Measurement commands are typically queries only and return power source measurement
data values. This section is broken down into the commands for each measurement
parameter.
The following measurement commands are supported.
8.4.1

Voltage Measurement Commands

Query Format
Description
Returned Data Format
Query Example
Query Format
Description
Returned Data Format
Query Example
Query Format
Description
Returned Data Format
Query Example
Note 1: Line to Line voltage measurements are calculated based on VLN and phase angles and are
valid only for sinusoidal voltage waveforms with low levels of distortion and under balanced three
phase load conditions.
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
LMX Series Power Source Operation Manual
MEASure:VOLTage[:ACDC]#?
Returns the measured RMS voltage for the selected phase #. If no
phase number is specified, returns the reading for the last selected
phase.
<nr2>
MEAS:VOLT:ACDC2?
230.0000
MEASure:VOLTage:AC#?
Returns the measured RMS voltage for the selected phase #. If no
phase number is specified, returns the reading for the last selected
phase. This command is equivalent to "MEASure:VOLTage[:ACDC]#?"
and is provided for backward compatibility with the UPC controllers.
<nr2>
MEAS:VOLT:AC3?
230.0000
MEASure:VLL#?
Returns the measured RMS Line-to-Line voltage
phase #. If no phase number is specified, returns the reading for the
last selected phase. Phase reference applies as follows:
# = 1 V
, # = 2 V
, # =3 V
AB
AC
MEAS:VLL2?
398.3780
LMX SERIES™ OPERATION MANUAL
Remote Control Programming
SECTION 8:
1
for the selected
BC
Page 208 of 463