6.7.2

USER LIMITS SETTINGS Screen

This screen allows programming user defined voltage and frequency upper and lower limits
to prevent an operator from accidentally programming output settings that could be
damaging to a unit under test.
For example, when testing a 50 Hz transformer, a lower frequency limit setting of 47 would
prevent output frequency programming of values that could cause the transformer to
saturate.
This applies to voltage where a high voltage value could damage a unit under test that was
not designed to handle high AC input voltage.
The following parameters can be set from this screen:
Voltage AC
Frequency
The soft keys on the USER LIMITS SETTINGS screen provide access to additional functionality
as listed in the table below.
SOFT KEY
Apply (SK1)
Cancel (SK2)
-
-
Back (SK5)
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
LMX Series Power Source Operation Manual

Figure 6-17: USER LIMIT SETTINGS Screen

Lower and Upper Vrms set limits for AC programming.
Lower and Upper limits for Frequency programming.
Description
Accepts new settings and returns to previous screen.
Returns to the previous screen.
Returns to the previous screen.