Pacific Power Source LMX Series Operation Manual page 390

Programmable ac power source
Table of Contents

Advertisement

10.7.3.3
PULSE Transients
PULSE transients allow series of pulsed output levels to be programmed without creating LIST
transients with repetitive entries. In PULSE transient mode, the user enters Level, duty cycle,
ramp time, period, initial value and pulse count. This mode is useful for high and low line testing
for voltage and frequency operating limits of an EUT.
The PULSE EXAMPLE shows the relationship between the various parameters.
This information is transformed to a standard Segment based transient list by the power source
when the LOAD button is pressed.
During transient execution, a progress indicator is visible showing percent completion.
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
LMX Series Power Source Operation Manual
LMX SERIES™ OPERATION MANUAL
LAN Interface Configuration
SECTION 10:
Page 390 of 463

Advertisement

Table of Contents
loading

Table of Contents