Query Syntax
Description
Returned Data Format
Query Example
Command Syntax
Description
Parameters
Parameter Format
Example
Query Format
Returned Data Format
Query Example
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
LMX Series Power Source Operation Manual
OUTPut:DISABLEPHase:MINimum
OUTPut:DISABLEPHase:MAXimum
OUTPut:DISABLEPHase:DEFault
These query commands return minimum, maximum and default
Output Disable Phase values respectively. Available range is -0.01
(Random) through 360.0.
Note: Available in units with Firmware revision 1.6.6 or higher.
<nr2>
OUTP:DISABLEPH:MIN?
-0.010
OUTP:DISABLEPH:MAX?
360.000
OUTP:DISABLEPH:DEF?
0.000
OUTPut:ZEROprogram
This command sets the programmed output voltage to zero before
opening the output relay when the OUTP OFF | 0 command is sent.
This feature is disabled by default but can be enabled be sending
OUTP:ZERO ON | 1.
In this mode, the output of the power source goes to a low
impedance state before disconnecting the load.
Note: Available in units with Firmware revision 1.6.6 or higher.
Refer also to the OUTPut:DISABLEPHase command.
< 0 | OFF | 1 | ON >
<b>
OUTP:ZERO 1
OUTPut:ZEROprogram?
<b>
OUTP:ZERO?
1
LMX SERIES™ OPERATION MANUAL
Remote Control Programming
SECTION 8: