Instrument Functions: A - L
FREQUENCY / Channel

2.7.3 Stop Freq

Sets the frequency at the right side of the graticule and sets the frequency entry mode to
Start/Stop. When the Start/Stop Frequency entry mode is activate, the start and stop
frequency values are displayed below the graticule in place of center frequency and span.
The left and right sides of the graticule correspond to the start and stop frequencies.
Key Path:
FREQUENCY
State Saved:
Saved in Instrument State
Factory Preset:
Model
E4440A
E4443A
E4445A
E4446A
E4448A
Range:
Model
E4440A
E4443A
E4445A
E4446A
E4448A
Remote Command:
[:SENSe]:FREQuency:STOP <frequency>
[:SENSe]:FREQuency:STOP?
Example:
FREQ:STOP 1600
FREQ:STOP?
146
Stop Frequency
26.50000000 GHz
6.70000000 GHz
13.20000000 GHz
44.00000000 GHz
50.00000000 GHz
Frequency Range
–100.0000000 MHz
–100.0000000 MHz
–100.0000000 MHz
–100.0000000 MHz
–100.0000000 MHz
to
27.0 GHz
to
13.7 GHz
to
7.2 GHz
to
44.5 GHz
to
51.0 GHz