Stop Frequency; Hardware Connection Instructions - Agilent Technologies PSA Series User Manual

Option 215 for the psa series of spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Source Key and Programming Commands
External Source Configuration

4.1.6 Stop Frequency

Information parameter which displays the stop frequency of the signal source connected at the specified IP
Address.
Mode:
Key Path:
Remote Command:
State Saved:
Dependencies and
Couplings:
Remote Command
Notes:
Example:

4.1.7 Hardware Connection Instructions

Pressing this immediate action keys displays a form showing you how to set up the connections between
the PSA and the external source. The form will be displayed until you press any key. See
Mode:
Key Path:
Dependencies and
Couplings:
Remote Command
Notes:
50
Spectrum Analysis
Source, Ext Src Config, Show Setup
:SOURce:EXTernal:FREQuency:STOP?
No
The external source stop frequency is changed each time the
stop frequency of the PSA changes.
If the calculated external source stop frequency exceeds the
frequency range of the external source Error Message 13227
will be reported and the sweep stopped until you adjust either
analyzer start or stop frequency, the frequency offset (if ON) or
harmonic (if ON) so that the frequency range can be swept by
the signal source and analyzer simultaneously.
Query only
SOUR:EXT:SWE:STOP?
Spectrum Analysis
Source, Ext Src Config
Parameter is disabled if active mode is not Spectrum Analyzer
mode.
No SCPI. Front Panel only.
Figure
4-2.