Weinmann OXYWAY Fix I Service Manual page 99

Pressure reducer
Table of Contents

Advertisement

174
173
174
Back
Forward
8. Remove the pressure reducer from the cover screw-in device 402.
9. Insert the pressure reducer into the cover clamping device 401.
When doing so, note: The cover clamping device 401 must clamp
onto the housing cover 162.
10.Tighten the cover clamping device 401 in the vise.
11.Push the counter cap 404 over the safety valve and the housing.
12.Place the torque wrench onto the square of the counter cap 404.
13.Screw the housing on tightly using the torque wrench.
404
14.Remove the torque wrench and counter cap 404.
15.Loosen the vise.
16.Remove the pressure reducer from the cover clamping device 401.
17.Remove the cover clamping device 401 from the vise.
14.14 Replacing the spring for the piston
14.14.1 Removing the spring for the piston
Requirement
The multistep flow selector assembly 179 has been removed
(see "14.13.1 Removing the multistep flow selector assembly",
page
1. Remove the spring 174 from the piston 173.
173
179
2. Remove the piston 173 from the multistep flow selector
assembly 179.
14.14.2 Fitting the spring for the piston
1. Position the spring 174 on the piston 173.
179
2. Insert the piston 173 with spring 174 into the multistep flow
selector assembly 179.
3. Fit the multistep flow selector assembly 179
(see "14.13.2 Fitting the multistep flow selector assembly",
page
Contents
Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
98).
98).
Previous page
OXYWAY
EN
99
Next page