Weinmann OXYWAY Fix I Service Manual page 72

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fix I, OXYWAY Fix III
409
72
EN
OXYWAY
Back
6. Place the torque wrench onto the square of the counter cap 404.
7. Use the torque wrench to loosen the housing from the connecting
404
113
8. Remove the torque wrench and counter cap 404.
9. Unscrew the housing from the connecting bolt 113 by hand and
10.Use the watchmaker's screwdriver 409 to remove the rib filter 104
104
113
13.10.2 Fitting the rib filter
1. Insert the rib filter 104 into the connecting bolt 113.
104
113
2. Screw the housing onto the connecting bolt 113 by hand.
3. Push the counter cap 404 over the safety valve and the housing.
4. Place the torque wrench onto the square of the counter cap 404.
5. Use the torque wrench to tightly screw the housing onto the
404
113
6. Remove the torque wrench and counter cap 404.
7. Loosen the hand-screwed connection 115 at the top part of
8. Unscrew the hand-screwed connection 115 from the bottom
9. Remove the hand-screwed connection 115 of the pressure
Forward
bolt 113.
then remove.
from the connecting bolt 113.
connecting bolt 113.
the counter tool 400.
part of the counter tool 400 by hand.
reducer from the counter tool 400.