Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
112
EN
OXYWAY
Back
• The multistep flow selector assembly 179 has been disassembled
1. Dispose of the sliding ring 164.
14.26.2 Fitting the sliding ring
1. Pick up the sliding ring 164.
2. Assemble the multistep flow selector assembly 179
3. Fit the piston 173
4. Fit the spring for the piston 174
5. Fit the multistep flow selector assembly 179
14.27 Disassembling/assembling the cover
assembly
Required tools
• Cover clamping device 401
• Vise
• Special counter tool, DISS 416
• Special gripping pliers 415
• Counter tool with fastening handle 400
• Counter cap 404
• Open-end wrench A/F 15
• Watchmaker's screwdriver 409
• Brush
• Oxygen-compatible lubricant 414
• Allen key size 5
• Thread tap G 1/8
• Brass brush
• Torque wrench 8 Nm ± 2 Nm
• Torque wrench 16 Nm ± 1 Nm
• Socket wrench size 15
• Loctite 413
Forward
(see "14.17.1 Disassembling the multistep flow selector assembly",
page
101).
(see "14.17.2 Assembling the multistep flow selector assembly",
page
102).
(see "14.16.2 Fitting the piston", page
(see "14.14.2 Fitting the spring for the piston", page
(see "14.13.2 Fitting the multistep flow selector assembly",
page
98).
Contents
Previous page
101).
99).
Next page