Weinmann OXYWAY Fix I Service Manual page 26

Pressure reducer
Table of Contents

Advertisement

Final inspection OXYWAY Fix I, OXYWAY Fix III
513
501
26
EN
OXYWAY
Back
8. Close the valve on the oxygen cylinder.
9. Slightly open the shut-off valve 508.
10.Close the shut-off valve 508.
11.Remove the shut-off valve 508 from the pressure outlet 105.
12.Remove the hand-screwed connection 115 from the valve of
7.7 Checking the static outlet
pressure p4
1. Screw the hand-screwed connection 115 onto the valve of
2. Screw the hand-screwed connection 115 tight by hand.
115
3. Screw the union nut of the supply test set 501 onto the pressure
4. Connect the thin hose of the supply test set 501 to the testing
Forward
Requirement:
The needle of the contents indicator 100 moves towards "0"
without catching.
Or
For pressure reducers with an output of 4 l/min:
• Remove the shut-off valve 508 from the adapter 507.
• Remove the adapter 507 from the pressure outlet 105.
the oxygen cylinder.
the oxygen cylinder.
outlet 105.
Or
For pressure reducers with an output of 4 l/min:
• Screw the adapter 507 onto the inhalation outlet 106.
• Screw the union nut of the supply test set 501 onto the
adapter 507.
device 513.
When doing so, note: The T-distributor with check valve is not
connected.