108
EN
OXYWAY
Back
14.21 Replacing the threaded pin in the

multistep flow selector

Required tools
Allen key size 2
14.21.1 Removing the threaded pin from the multistep
flow selector
Requirement
• The multistep flow selector assembly 179 has been removed
• The spring for the piston 174 has been removed
• The piston 173 has been removed
• The multistep flow selector assembly 179 has been disassembled
1. Use the Allen key to remove the threaded pin 160 from the
159
160
14.21.2 Fitting the threaded pin in the multistep flow
selector
1. Insert the threaded pin 160 into the multistep flow selector 159
159
160
2. Assemble the multistep flow selector assembly 179
3. Fit the piston 173
4. Fit the spring for the piston 174
5. Fit the multistep flow selector assembly 179
Forward
(see "14.13.1 Removing the multistep flow selector assembly",
page
98).
(see "14.14.1 Removing the spring for the piston", page
piston", page
100).
(see "14.17.1 Disassembling the multistep flow selector assembly",
page
101).
multistep flow selector 159.
and screw in using the Allen key.
When doing so, note: The threaded pin 160 must not be screwed
into the hole for the orifice plate assembly 181.
(see "14.17.2 Assembling the multistep flow selector assembly",
page
102).
(see "14.16.2 Fitting the piston", page
(see "14.14.2 Fitting the spring for the piston", page
(see "14.13.2 Fitting the multistep flow selector assembly",
page
98).
Contents
Previous page
99).
(see "14.16.1 Removing the
101).
99).
Next page

Advertisement