Weinmann OXYWAY Fix I Service Manual page 152

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fast II
152
EN
OXYWAY
Back
15.21.2 Fitting the O-ring in the contents indicator
1. Position the O-ring 202 on the contents indicator 200.
2. Fit the contents indicator 200
3. Fit the threaded locking pin 201
4. Fit the multistep flow selector assembly 233
15.22 Replacing the contents indicator
Required tools
• Oxygen-compatible lubricant 414
• Brush
15.22.1 Removing the contents indicator
Requirement
• The multistep flow selector assembly 233 has been removed
• The threaded locking pin 201 has been removed
1. Remove the contents indicator 200 from the housing.
200
2. Remove the O-ring 202 from the contents indicator 200.
15.22.2 Fitting the contents indicator
1. Use the brush to wet the groove in the contents indicator 200
2. Position the O-ring 202 on the contents indicator 200.
3. Insert the contents indicator 200 into the housing.
200
Forward
(see "15.22.2 Fitting the contents indicator", page
(see "15.18.2 Fitting the threaded locking pin", page
(see "15.16.2 Fitting the multistep flow selector assembly",
page
145).
(see "15.16.1 Removing the multistep flow selector assembly",
page
144).
(see "15.18.1 Removing the threaded locking pin", page
with oxygen-compatible lubricant 414.
Contents
Previous page
152).
147).
147).
Next page