Weinmann OXYWAY Fix I Service Manual page 222

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Click
318
307
317
307
222
EN
OXYWAY
Back
17.17.2 Assembling the cover assembly
1. Insert the large O-ring 318 into the groove in the multistep flow
2. Insert the small O-ring 317 into the groove in the multistep flow
3. Use the Allen key to insert the O-ring 319 into the groove in the
308
319
4. Use the brush to apply a thin coat of oxygen-compatible
5. Place the housing cover 308 in the cover clamping device 401.
6. Tighten the cover clamping device 401 in the vise.
7. Place the multistep flow selector 307 on the housing cover 308.
8. Pull the union nut 316 over the multistep flow selector 307 and
316
307
308
9. Remove the adhesive residue from the threads on the multistep
Forward
selector 307 by hand.
selector 307 by hand.
housing cover 308.
lubricant 414 to the O-ring 319 in the housing cover 308.
onto the housing cover 308 and screw on by hand.
flow selector 307.
When doing so, note:
• No dirt must get into the housing during cleaning.
• A brass brush must be used for the external thread.
• A thread tap must be used for the internal thread.