13.9 Replacing the hand-screwed
connection
13.9.1 Removing the hand-screwed connection
Requirement
• The O-ring 114 has been removed from the connecting bolt 113
(see "13.6.1 Removing the O-ring from the connecting bolt",
page
• The connecting bolt 113 has been removed
(see "13.8.1 Removing the connecting bolt", page
1. Dispose of the hand-screwed connection 115.
13.9.2 Fitting the hand-screwed connection
1. Pick up the hand-screwed connection 115.
2. Fit the connecting bolt 113
(see "13.8.2 Fitting the connecting bolt", page
3. Fit the O-ring 114 in the connecting bolt 113
(see "13.6.2 Fitting the O-ring in the connecting bolt", page
13.10 Replacing the rib filter
Required tools
• Counter tool with fastening handle 400
• Counter cap 404
• Vise
• Torque wrench 60 Nm ± 1 Nm
• Watchmaker's screwdriver 409
13.10.1 Removing the rib filter
Requirement
The O-ring 114 has been removed from the connecting bolt 113
(see "13.6.1 Removing the O-ring from the connecting bolt",
page
68).
1. Clamp the bottom part of the counter tool 400 in the vise.
2. Put the top part of the counter tool 400 onto the bottom part of
the counter tool 400.
3. Screw the hand-screwed connection 115 of the pressure reducer
onto the bottom part of the counter tool 400 by hand.
4. Tightly screw the hand-screwed connection 115 onto the top part
of the counter tool 400.
5. Push the counter cap 404 over the safety valve and the housing.
Contents