Repairs to OXYWAY Fine I
186
EN
OXYWAY
Back
7. Remove any adhesive residue from the threads on the housing.
16.13.2 Fitting the inhalation outlet
1. Pick up the inhalation outlet 257.
Risk due to adhesive in the gas flow!
If the first two threads of the inhalation outlet are moistened with
adhesive, this adhesive can get into the gas flow to the patient and
lead to airway irritation.
 Do not moisten the first two threads with adhesive.
2. Apply Loctite 413 to the thread of the inhalation outlet 257.
3. Insert the inhalation outlet 257 into the housing.
257
4. Use the torque wrench to tightly screw in the inhalation outlet 257
5. Loosen the hand-screwed connection 273 at the top part of the
6. Unscrew the hand-screwed connection 273 from the bottom part
7. Remove the hand-screwed connection 273 of the pressure reducer
8. Remove the top part of the counter tool 400 from the bottom part
9. Loosen the bottom part of the counter tool 400 from the vise and
10.Fit the sealing disks 260 and nozzle plate 259 in the inhalation
11.Fit the pressure screw 258 in the inhalation outlet 257
12.Fit the O-ring 272 in the connecting bolt 271
Forward
When doing so, note:
• No dirt must get into the housing during cleaning.
• A brass brush must be used for the external thread.
• A thread tap must be used for the internal thread.
in the housing.
When doing so, note: Loctite 413 must harden according to the
manufacturer's specifications.
counter tool 400.
of the counter tool 400 by hand.
from the counter tool 400.
of the counter tool 400.
remove.
outlet 257
(see "16.6.2 Fitting the sealing disks and nozzle plate in
the inhalation outlet", page
(see "16.5.2 Fitting the pressure screw in the inhalation outlet",
page
178).
(see "16.8.2 Fitting the O-ring in the connecting bolt", page
Contents
Previous page
179).
Next page
180).