Weinmann OXYWAY Fix I Service Manual page 34

Pressure reducer
Table of Contents

Advertisement

Final inspection OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
513
501
34
EN
OXYWAY
Back
4. Screw the adapter 507 onto the inhalation outlet 155.
5. Screw the union nut of the supply test set 501 onto the
6. Connect the thin hose of the supply test set 501 to the testing
7. Open the valve of the oxygen cylinder slowly.
8. Wait 1 min until the static outlet pressure at the test equipment
9. Set the multistep flow selector 159 to "15".
10.Read off the cylinder pressure p1 at the contents indicator 150
11.Enter the static outlet pressure p4 at the testing device 513 and
12.Close the valve on the oxygen cylinder.
13.Remove the thin hose of the supply test set 501 from the testing
14.Remove the union nut of the supply test set 501 from the
15.Remove the adapter 507 from the inhalation outlet 155.
16.Remove the hand-screwed connection 177 from the valve of
Forward
adapter 507.
device 513.
When doing so, note: The T-distributor with check valve is not
connected.
has adjusted itself and remains constant.
and enter the value in the test report.
enter the value in the test report.
Requirement:
The static test pressure p4 is within the tolerance range according
to the following table:
Cylinder pressure p1
(in bar)
20
80
100
140
200
Requirement:
The static outlet pressure p4 remains constant at the previously
measured value after 1 minute.
device 513.
adapter 507.
the oxygen cylinder.
Contents
Previous page
Static outlet pressure p4
(in bar)
4.3 ± 0.4
4.6 ± 0.4
4.7 ± 0.4
4.8 ± 0.4
4.9 ± 0.5
Next page