1. Remove the driving plate 266 from the square in the spindle 270 by
266
hand.
16.17.2 Fitting the driving plate
1. Place the driving plate 266 on the square in the spindle 270 by
266
hand.
2. Fit the multistep flow selector 267
(see "16.16.2 Fitting the multistep flow selector", page
3. Fit the hexagon nut 268
(see "16.15.2 Fitting the hexagon nut", page
4. Fit the cover for the multistep flow selector 269
(see "16.14.2 Fitting the cover for the multistep flow selector",
page
16.18 Replacing the assembly housing
cover
Required tools
• Cover clamping device 401
• Cover screw-in device 402
• Counter cap 404
• Torque wrench 8 Nm ± 2 Nm
• Vise
16.18.1 Removing the assembly housing cover
Requirement
• The cover for the multistep flow selector 269 has been removed
(see "16.14.1 Removing the cover for the multistep flow selector",
page
• The hexagon nut 268 has been removed
(see "16.15.1 Removing the hexagon nut", page
• The multistep flow selector 267 has been removed
(see "16.16.1 Removing the multistep flow selector", page
• The driving plate 266 has been removed
(see "16.17.1 Removing the driving plate", page
Contents
187).
187).
Previous page
Repairs to OXYWAY Fine I
188).
188).
187).
188).
188).