Weinmann OXYWAY Fix I Service Manual page 143

Pressure reducer
Table of Contents

Advertisement

Back
Forward
15.15.1 Removing the O
Requirement
The O-ring 229 has been removed from the connecting bolt 228
(see "15.8.1 Removing the O-ring from the connecting bolt",
page
134).
1. Clamp the bottom part of the counter tool 400 in the vise.
2. Put the top part of the counter tool 400 onto the bottom part of
the counter tool 400.
3. Screw the hand-screwed connection 230 of the pressure reducer
onto the bottom part of the counter tool 400 by hand.
4. Tightly screw the hand-screwed connection 230 onto the top part
of the counter tool 400.
5. Use the open-end wrench to loosen the O
reducer from the housing.
6. Remove the O
232
7. Remove any adhesive residue from the threads on the housing.
When doing so, note:
• No dirt must get into the housing during cleaning.
• A brass brush must be used for the external thread.
• A thread tap must be used for the internal thread.
15.15.2 Fitting the O
1. Pick up the O
Risk due to adhesive in the gas flow!
If the first two threads of the O
adhesive, this adhesive can get into the gas flow to the patient
and lead to airway irritation.
 Do not moisten the first two threads with adhesive.
2. Apply Loctite 413 to the thread of the O
3. Tightly screw the reducer into the O
When doing so, note: Loctite 413 must harden according to
the manufacturer's specifications.
4. Apply Loctite 413 to the thread of the reducer.
Contents
coupling
2
coupling 232 and reducer from the housing.
2
coupling
2
coupling 232.
2
coupling are moistened with
2
coupling 232 by hand.
2
Previous page
Repairs to OXYWAY Fast II
coupling 232 and
2
coupling 232.
2
OXYWAY
EN
143