Weinmann OXYWAY Fix I Service Manual page 134

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fast II
228
409
228
229
134
EN
OXYWAY
Back
15.7.2 Fitting the pressure screw in the pressure outlet
1. Insert the pressure screw 210 into the pressure outlet 209.
209
210
2. Use the slot-head screwdriver to tighten the pressure screw 210
15.8 Replacing the O-ring in the
connecting bolt
Required tools
Watchmaker's screwdriver 409
15.8.1 Removing the O-ring from the connecting bolt
1. Use the watchmaker's screwdriver 409 to carefully lift up the
229
15.8.2 Fitting the O-ring in the connecting bolt
1. If the groove in the connecting bolt 228 is soiled:
2. Insert the O-ring 229 into the groove in the connecting bolt 228
Forward
in the pressure outlet 209.
O-ring 229 and remove it from the groove in the connecting
bolt 228.
When doing so, note: The groove must not be damaged.
Clean the groove in the connecting bolt 228 with a dry cloth.
by hand.
Contents
Previous page
Next page

Advertisement

Table of Contents
loading

Table of Contents