17.20.2 Fitting the housing cover
1. Pick up the housing cover 308.
2. Fit the O-ring 319 in the housing cover 308
(see "17.18.2 Fitting the O-ring in the housing cover", page
3. Assemble the cover assembly 324
(see "17.17.2 Assembling the cover assembly", page
4. Assemble the multistep flow selector assembly 323
(see "17.9.2 Assembling the multistep flow selector assembly",
page
213).
5. Fit the multistep flow selector assembly 323
(see "17.8.2 Fitting the multistep flow selector assembly",
page
211).
6. Attach the label 321
17.21 Replacing the multistep flow selector
17.21.1 Removing the multistep flow selector
Requirement
• The multistep flow selector assembly 323 has been removed
(see "17.8.1 Removing the multistep flow selector assembly",
page
211).
• The multistep flow selector assembly 323 has been disassembled
(see "17.9.1 Disassembling the multistep flow selector assembly",
page
212).
• The cover assembly 324 has been disassembled
(see "17.17.1 Disassembling the cover assembly", page
1. Dispose of the multistep flow selector 307.
17.21.2 Fitting the multistep flow selector
1. Pick up the multistep flow selector 307.
2. Assemble the cover assembly 324
(see "17.17.2 Assembling the cover assembly", page
3. Assemble the multistep flow selector assembly 323
(see "17.9.2 Assembling the multistep flow selector assembly",
page
213).
4. Fit the multistep flow selector assembly 323
(see "17.8.2 Fitting the multistep flow selector assembly",
page
211).
Contents
Repairs to OXYWAY Click
(see "17.4.2 Attaching a label", page
OXYWAY