Weinmann OXYWAY Fix I Service Manual page 100

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
100
EN
OXYWAY
Back
14.15 Replacing the threaded locking pin
Required tools
Allen key size 1.5
14.15.1 Removing the threaded locking pin
Requirement
The multistep flow selector assembly 179 has been removed
(see "14.13.1 Removing the multistep flow selector assembly",
page
1. Use the Allen key to loosen the threaded locking pin 151 from
2. Remove the threaded locking pin 151 from the housing.
151
14.15.2 Fitting the threaded locking pin
1. Insert the threaded locking pin 151 into the housing.
151
2. Use the Allen key to lightly screw in the threaded locking pin 151
3. Fit the multistep flow selector assembly 179
14.16 Replacing the piston
14.16.1 Removing the piston
Requirement
• The multistep flow selector assembly 179 has been removed
• The spring for the piston 174 has been removed
1. Remove the piston 173 from the multistep flow selector
Forward
98).
the housing.
in the housing.
the multistep flow selector assembly", page
(see "14.13.1 Removing the multistep flow selector assembly",
page
98).
(see "14.14.1 Removing the spring for the piston", page
assembly 179.
Contents
Previous page
(see "14.13.2 Fitting
98).
99).
Next page