Weinmann OXYWAY Fix I Service Manual page 191

Pressure reducer
Table of Contents

Advertisement

262
Back
Forward
12.Loosen the vise.
13.Remove the pressure reducer from the cover clamping device 401.
14.Remove the cover clamping device 401 from the vise.
15.Fit the driving plate 266
(see "16.17.2 Fitting the driving plate", page
16.Fit the multistep flow selector 267
(see "16.16.2 Fitting the multistep flow selector", page
17.Fit the hexagon nut 268
(see "16.15.2 Fitting the hexagon nut", page
18.Fit the cover for the multistep flow selector 269
(see "16.14.2 Fitting the cover for the multistep flow selector",
page
16.19 Replacing the spring for the piston
16.19.1 Removing the spring for the piston
Requirement
• The cover for the multistep flow selector 269 has been removed
(see "16.14.1 Removing the cover for the multistep flow selector",
page
• The hexagon nut 268 has been removed
(see "16.15.1 Removing the hexagon nut", page
• The multistep flow selector 267 has been removed
(see "16.16.1 Removing the multistep flow selector", page
• The driving plate 266 has been removed
(see "16.17.1 Removing the driving plate", page
• The assembly housing cover 275 has been removed
(see "16.18.1 Removing the assembly housing cover", page
1. Remove the spring 262 from the piston 261.
261
2. Remove the piston 261 from the assembly housing cover 275.
Contents
187).
187).
Previous page
Repairs to OXYWAY Fine I
189).
188).
188).
187).
188).
188).
189).
OXYWAY
EN
191
Next page