Weinmann OXYWAY Fix I Service Manual page 27

Pressure reducer
Table of Contents

Advertisement

Back
Forward
5. Open the valve of the oxygen cylinder slowly.
6. Wait 1 min until the static outlet pressure at the test equipment has
adjusted itself and remains constant.
7. Read off the cylinder pressure p1 at the contents indicator 100
and enter the value in the test report.
8. Enter the static outlet pressure p4 at the testing device 513
and enter the value in the test report.
Requirement:
The static outlet pressure p4 is within the tolerance range according
to the following table:
Cylinder pressure p1
(in bar)
20
80
100
140
200
Requirement:
The static outlet pressure p4 remains constant at the previously
measured value after 1 minute.
9. Close the valve on the oxygen cylinder.
10.Remove the thin hose of the supply test set 501 from the testing
device 513.
11.Remove the union nut of the supply test set 501 from the pressure
outlet 105.
Or
For pressure reducers with an output of 4 l/min:
• Remove the union nut of the supply test set 501 from
the adapter 507.
• Remove the adapter 507 from the inhalation outlet 106.
12.Remove the hand-screwed connection 115 from the valve of
the oxygen cylinder.
Contents
Final inspection OXYWAY Fix I, OXYWAY Fix III
Static outlet pressure p4
(in bar)
4.3 ± 0.3
4.5 ± 0.3
4.6 ± 0.3
4.7 ± 0.3
4.9 ± 0.5
OXYWAY
Previous page
EN
27