Weinmann OXYWAY Fix I Service Manual page 28

Pressure reducer
Table of Contents

Advertisement

Final inspection OXYWAY Fix I, OXYWAY Fix III
510/511
500
Cylinder pressure
OXYWAY Fix III:
p1 (in bar)
Pressure reducer
with an output of
Q1:190 l/min
100
> 160 l/min
28
EN
OXYWAY
Back
7.8 Checking the constant output (flow)
1. Screw the hand-screwed connection 115 onto the valve of
2. Screw the hand-screwed connection 115 tight by hand.
115
3. Screw the special kit 500 onto the pressure outlet 105.
4. Connect the special kit 500 to the testing device 511.
5. Open the valve of the oxygen cylinder slowly.
6. Read off the cylinder pressure p1 at the contents indicator 100
7. Read off the flow values at the testing device 510 or 511 and
OXYWAY Fix III:
Pressure reducer with DIN or
AGA coupling with an output
of Q1:150 l/min
> 120 l/min
8. Close the valve on the oxygen cylinder.
9. Remove the special kit 500 from the testing device 510 or 511.
10.Remove the special kit 500 from the pressure outlet 105.
11.Remove the hand-screwed connection 115 from the valve of
Forward
the oxygen cylinder.
Or
For pressure reducers with an output of 4 l/min:
• Screw the adapter 507 onto the inhalation outlet 106.
• Screw the special kit 500 onto the adapter 507.
• Connect the special kit 500 to the testing device 510.
and enter the value in the test report.
enter in the test report.
Requirement:
The flow value is within the tolerance range according to
the following table:
OXYWAY Fix III:
Pressure reducer with
bayonet coupling
> 120 l/min
Or
For pressure reducers with an output of 4 l/min:
• Remove the special kit 500 from the adapter 507
• Remove the adapter 507 from the inhalation outlet 106.
the oxygen cylinder.
Contents
Previous page
OXYWAY Fix I:
Pressure reducer
with an output of
Q1:4 l/min
4
± 0.5
Next page