Weinmann OXYWAY Fix I Service Manual page 59

Pressure reducer
Table of Contents

Advertisement

12 Faults
Fault
Output value outside the specifications
Leakage at the connecting bolt
(cylinder connection)
(exception: OXYWAY Click)
Back
If you are not able to rectify a fault with the aid of the table below,
please contact the manufacturer WEINMANN Emergency. To avoid
serious damage, do not allow your customer to continue using the
device.
Cause
Disposable filter soiled
O-ring damaged
Connecting bolt loose
Connecting bolt damaged
Forward
Remedy
Replace the disposable filter
• OXYWAY Fix I, OXYWAY Fix III:
(see "13.7 Replacing the disposable filter", page 68)
• OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III:
(see "14.8 Replacing the disposable filter", page 91)
• OXYWAY Fast II:
(see "15.9 Replacing the disposable filter", page 135)
• OXYWAY Fine I:
(see "16.9 Replacing the disposable filter", page 180)
• OXYWAY Click:
(see "17.28 Replacing the disposable filter", page 233)
Replace the O-ring
• OXYWAY Fix I, OXYWAY Fix III:
(see "13.6 Replacing the O-ring in the connecting bolt",
page 68)
• OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III:
(see "14.7 Replacing the O-ring in the connecting bolt",
page 90)
• OXYWAY Fast II:
(see "15.8 Replacing the O-ring in the connecting bolt",
page 134)
• OXYWAY Fine I:
(see "16.8 Replacing the O-ring in the connecting bolt",
page 179)
Tighten the connecting bolt
• OXYWAY Fix I, OXYWAY Fix III:
(see "13.8.2 Fitting the connecting bolt", page 70)
• OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III:
(see "14.9.2 Fitting the connecting bolt", page 93)
• OXYWAY Fast II:
(see "15.10.2 Fitting the connecting bolt", page 136)
• OXYWAY Fine I:
(see "16.10.2 Fitting the connecting bolt", page 182)
Replace the connecting bolt
• OXYWAY Fix I, OXYWAY Fix III:
(see "13.8 Replacing the connecting bolt", page 69)
• OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III:
(see "14.9 Replacing the connecting bolt", page 92)
• OXYWAY Fast II:
(see "15.10 Replacing the connecting bolt", page 135)
• OXYWAY Fine I:
(see "16.10 Replacing the connecting bolt", page 181)
Contents
Previous page
Faults
OXYWAY
EN
59
Next page

Advertisement

Table of Contents
loading

Table of Contents