Weinmann OXYWAY Fix I Service Manual page 60

Pressure reducer
Table of Contents

Advertisement

Faults
Fault
Leakage at the pressure outlet/
inhalation outlet
Bent contents indicator
(exception: OXYWAY Click)
Needle zero point deviation on
the contents indicator
(exception: OXYWAY Click)
Bent connection on the contents indicator
(exception: OXYWAY Click)
Bent flow indicator
(OXYWAY Fine I only)
Needle zero point deviation on
the flow indicator
(OXYWAY Fine I only)
Bent connection on the flow indicator
(OXYWAY Fine I only)
60
EN
OXYWAY
Back
Cause
Pressure outlet/inhalation outlet
loose
Pressure outlet/inhalation outlet
damaged
Contents indicator defective
Flow indicator defective
Forward
Remedy
Tighten the pressure outlet/inhalation outlet
• OXYWAY Fix I, OXYWAY Fix III:
(see "13.11.2 Fitting the pressure outlet/inhalation outlet",
page 74)
• OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III:
(see "14.12.2 Fitting the inhalation outlet", page 96)
• OXYWAY Fast II:
(see "15.13.2 Fitting the inhalation outlet", page 140)
(see "15.14.2 Fitting the pressure outlet", page 142)
• OXYWAY Fine I:
(see "16.13.2 Fitting the inhalation outlet", page 186)
• OXYWAY Click:
(see "17.7.2 Fitting the inhalation outlet", page 210)
Replace the pressure outlet/inhalation outlet
• OXYWAY Fix I, OXYWAY Fix III:
(see "13.11 Replacing the pressure outlet/inhalation outlet",
page 73)
• OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III:
(see "14.12 Replacing the inhalation outlet", page 95)
• OXYWAY Fast II:
(see "15.13 Replacing the inhalation outlet", page 139)
(see "15.14 Replacing the pressure outlet", page 141)
• OXYWAY Fine I:
(see "16.13 Replacing the inhalation outlet", page 185)
• OXYWAY Click:
(see "17.7 Replacing the inhalation outlet", page 209)
Replace the contents indicator
• OXYWAY Fix I, OXYWAY Fix III:
(see "13.16 Replacing the contents indicator", page 79)
• OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III:
(see "14.19 Replacing the contents indicator", page 105)
• OXYWAY Fast II:
(see "15.22 Replacing the contents indicator", page 152)
• OXYWAY Fine I:
(see "16.25 Replacing the contents indicator", page 198)
Replace the flow indicator
• OXYWAY Fine I:
(see "16.26 Replacing the flow indicator", page 200)
Contents
Previous page
Next page

Advertisement

Table of Contents
loading