Repairs to OXYWAY Fix I, OXYWAY Fix III
76
EN
OXYWAY
Back
13.12.2 Fitting the O
1. Pick up the O
Risk due to adhesive in the gas flow!
If the first two threads of the O
adhesive, this adhesive can get into the gas flow to the patient
and lead to airway irritation.
 Do not moisten the first two threads with adhesive.
2. Apply Loctite 413 to the thread of the O
3. Tightly screw the reducer onto the O
4. Apply Loctite 413 to the thread of the reducer.
5. Insert the O
117
6. Use the torque wrench to tightly screw in the O
7. Loosen the hand-screwed connection 115 at the top part of the
8. Unscrew the hand-screwed connection 115 from the bottom part
9. Remove the hand-screwed connection 115 of the pressure reducer
10.Remove the top part of the counter tool 400 from the bottom part
11.Loosen the bottom part of the counter tool 400 from the vise and
12.Fit the O-ring 114 in the connecting bolt 113
13.13 Replacing the spring for the piston
Required tools
• Cover clamping device 401
• Cover screw-in device 402
• Counter cap 404
• Torque wrench 8 Nm ± 2 Nm
• Vise
Forward
coupling
2
coupling.
2
When doing so, note: Loctite 413 must harden according to
the manufacturer's specifications.
coupling 117 and reducer into the housing.
2
reducer in the housing.
When doing so, note: Loctite 413 must harden according to the
manufacturer's specifications.
counter tool 400.
of the counter tool 400 by hand.
from the counter tool 400.
of the counter tool 400.
remove.
(see "13.6.2 Fitting the O-ring in the connecting bolt", page
Contents
Previous page
coupling are moistened with
2
coupling 117.
2
coupling 117 by hand.
2
coupling 117 and
2
Next page
68).