Weinmann OXYWAY Fix I Service Manual page 142

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fast II
142
EN
OXYWAY
Back
15.14.2 Fitting the pressure outlet
1. Pick up the pressure outlet 209.
Risk due to adhesive in the gas flow!
If the first two threads of the pressure outlet are moistened with
adhesive, this adhesive can get into the gas flow to the patient
and lead to airway irritation.
 Do not moisten the first two threads with adhesive.
2. Apply Loctite 413 to the thread of the pressure outlet 209.
3. Insert the pressure outlet 209 into the housing.
209
4. Use the torque wrench to tightly screw the pressure outlet 209
5. Loosen the hand-screwed connection 230 at the top part of the
6. Unscrew the hand-screwed connection 230 from the bottom part
7. Remove the hand-screwed connection 230 of the pressure reducer
8. Remove the top part of the counter tool 400 from the bottom part
9. Loosen the bottom part of the counter tool 400 from the vise and
10.Fit the pressure screw 210 in the pressure outlet 209
11.Fit the O-ring 229 in the connecting bolt 228
15.15 Replacing the O
Required tools
• Counter tool with fastening handle 400
• Vise
• Loctite 413
• Open-end wrench A/F 19
• Torque wrench 16 Nm ± 1 Nm
• Brass brush
• Thread tap G 1/8
Forward
in the housing.
When doing so, note: Loctite 413 must harden according to the
manufacturer's specifications.
counter tool 400.
of the counter tool 400 by hand.
from the counter tool 400.
of the counter tool 400.
remove.
(see "15.7.2 Fitting the pressure screw in the pressure outlet",
page
134).
(see "15.8.2 Fitting the O-ring in the connecting bolt", page
Contents
Previous page
coupling
2
Next page
134).

Advertisement