Maintenance
5 Maintenance
18
EN
OXYWAY
Back
Fault or treatment failure due to insufficient or incorrect
maintenance!
Incorrect maintenance can lead to dangerous situations and failure
or malfunctioning of the device. This can result in severe or life-
threatening injury to the patient and bystanders.
 Make sure that maintenance, technical safety checks and
 Also observe and comply with the maintenance intervals for
 Follow the maintenance intervals as stated on the device labeling.
 Perform a full function check before every use.
5.1 General instructions
• A final inspection must be performed after every maintenance
• Maintenance, technical safety checks, inspections and repairs
5.2 Intervals
Part concerned Interval
Pressure reducer
Disposable filter
(exceptions:
OXYWAY Click and
pressure reducers
with PIN-Index
connection)
Forward
maintenance measures are only performed by the manufacturer
or technicians who have been expressly authorized by the
manufacturer.
devices in storage.
procedure or technical safety check.
must only be carried out by WEINMANN Emergency or a technician
specifically authorized by WEINMANN Emergency.
Maintenance every
6 years
Technical Safety Check
("Sicherheitstechnische
Kontrolle" STK* in
accordance with § 11 of
the German regulation
MPBetreibV) every 2 years
Replace every 2 years
(or immediately if soiled)
Contents
Previous page
Maintenance by
Manufacturer or a technician
specifically authorized by the
manufacturer
User/operator
• OXYWAY Fix I,
OXYWAY Fix III:
(see "13.7 Replacing the
disposable filter", page 68)
• OXYWAY Fast I,
OXYWAY Fast I+,
OXYWAY Fast III:
(see "14.8 Replacing the
disposable filter", page 91)
• OXYWAY Fast II:
(see "15.9 Replacing the
disposable filter", page 135)
• OXYWAY Fine I:
(see "16.9 Replacing the
disposable filter", page 180)
Next page