Weinmann OXYWAY Fix I Service Manual page 187

Pressure reducer
Table of Contents

Advertisement

267
267
Back
Forward
16.14 Replacing the cover for the multistep
flow selector
Required tools
Knife
16.14.1 Removing the cover for the multistep flow
selector
1. Slide the blade of the knife under the cover 269.
267
2. Use the blade of the knife to detach the cover 269 from the
269
multistep flow selector 267 and then remove it.
16.14.2 Fitting the cover for the multistep flow selector
1. Insert the cover 269 into the recess in the multistep flow
selector 267 by hand, and press it down.
269
16.15 Replacing the hexagon nut
Required tools
Socket wrench size 8
16.15.1 Removing the hexagon nut
Requirement
The cover for the multistep flow selector 269 has been removed
(see "16.14.1 Removing the cover for the multistep flow selector",
page
187).
1. Turn the multistep flow selector 267 counter-clockwise to the stop.
2. Remove the hexagon nut 268 using the socket wrench.
When doing so, note: Hold the multistep flow selector 267 with
268
your hand during this step.
Contents
Repairs to OXYWAY Fine I
OXYWAY
Previous page
EN
187
Next page

Advertisement

Table of Contents
loading