Repairs to OXYWAY Fix I, OXYWAY Fix III
80
EN
OXYWAY
Back
1. Remove the contents indicator 100 from the housing.
100
2. Remove the O-ring 102 from the contents indicator 100.
13.16.2 Fitting the contents indicator
1. Use the brush to wet the groove in the contents indicator 100
2. Position the O-ring 102 on the contents indicator 100.
3. Insert the contents indicator 100 into the housing.
100
4. Fit the threaded locking pin 101
5. Fit the spring for the piston 111
13.17 Replacing the O-ring in the contents
indicator
Required tools
Watchmaker's screwdriver 409
13.17.1 Removing the O-ring from the contents
indicator
Requirement
• The spring for the piston 111 has been removed
• The threaded locking pin 101 has been removed
• The contents indicator 100 has been removed
1. Use the watchmaker's screwdriver 409 to carefully remove
Forward
with oxygen-compatible lubricant 414.
(see "13.15.2 Fitting the threaded locking pin", page
(see "13.13.2 Fitting the spring for the piston", page
(see "13.13.1 Removing the spring for the piston", page
(see "13.15.1 Removing the threaded locking pin", page
(see "13.16.1 Removing the contents indicator", page
the O-ring 102 from the housing.
Contents
Previous page
79).
77).
77).
79).
79).