Weinmann OXYWAY Fix I Service Manual page 221

Pressure reducer
Table of Contents

Advertisement

415
316
317
409
307
318
409
307
Back
Forward
5. Screw the special counter tool DISS 416 onto the inhalation
301
outlet 301.
416
6. Position the special gripping pliers 415 on the union nut 316.
7. Use the special gripping pliers 415 and the special counter tool
DISS 416 to slightly loosen the union nut 316 from the housing
cover 308.
308
When doing so, note: Do not yet remove the union nut 316 from
the housing cover 308.
8. Remove the special counter tool DISS 416 from the inhalation
outlet 301.
9. Use the open-end wrench to remove the inhalation outlet 301 from
the multistep flow selector 307.
10.Remove the housing cover 308 from the cover clamping device 401
together with the union nut 316 and the multistep flow selector
307.
11.Unscrew the union nut 316 from the housing cover 308 by hand.
12.Remove the union nut 316.
316
13.Remove the multistep flow selector 307.
307
308
14.Use the watchmaker's screwdriver 409 to carefully lift the
O-ring 319 out of the housing cover 308.
308
319
S
15.Use the watchmaker's screwdriver 409 to carefully lift the small
O-ring 317 out of the multistep flow selector 307.
16.Use the watchmaker's screwdriver 409 to carefully lift the large
O-ring 318 out of the multistep flow selector 307.
Contents
Repairs to OXYWAY Click
OXYWAY
Previous page
EN
221
Next page