16.13 Replacing the inhalation outlet
Required tools
• Counter tool with fastening handle 400
• Open-end wrench A/F 15
• Vise
• Brass brush
• Loctite 413
• Torque wrench 8 Nm ± 2 Nm
• Thread tap G 1/8
16.13.1 Removing the inhalation outlet
Requirement
• The O-ring 272 has been removed from the connecting bolt 271
(see "16.8.1 Removing the O-ring from the connecting bolt",
page
• The pressure screw 258 has been removed from the inhalation
outlet 257
inhalation outlet", page
• The sealing disks 260 and the nozzle plate 259 have been removed
from the inhalation outlet 257
disks and nozzle plate from the inhalation outlet", page
1. Clamp the bottom part of the counter tool 400 in the vise.
2. Put the top part of the counter tool 400 onto the bottom part of the
counter tool 400.
3. Screw the hand-screwed connection 273 of the pressure reducer
onto the bottom part of the counter tool 400 by hand.
4. Tightly screw the hand-screwed connection 273 onto the top part
of the counter tool 400.
5. Use the open-end wench to loosen the inhalation outlet 257 from
the housing.
6. Remove the inhalation outlet 257 from the housing.