14.18.2 Fitting the O-ring in the contents indicator
1. Position the O-ring 152 on the contents indicator 150.
2. Fit the contents indicator 150
(see "14.19.2 Fitting the contents indicator", page
3. Fit the threaded locking pin 151
(see "14.15.2 Fitting the threaded locking pin", page
4. Fit the multistep flow selector assembly 179
the multistep flow selector assembly", page
14.19 Replacing the contents indicator
Required tools
• Oxygen-compatible lubricant 414
• Brush
14.19.1 Removing the contents indicator
Requirement
• The multistep flow selector assembly 179 has been removed
(see "14.13.1 Removing the multistep flow selector assembly",
page
• The threaded locking pin 151 has been removed
(see "14.15.1 Removing the threaded locking pin", page
1. Remove the contents indicator 150 from the housing.
150
2. Remove the O-ring 152 from the contents indicator 150.
14.19.2 Fitting the contents indicator
1. Use the brush to wet the groove in the contents indicator 150
with oxygen-compatible lubricant 414.
2. Position the O-ring 152 on the contents indicator 150.
3. Insert the contents indicator 150 into the housing.
150
Contents
Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
98).
Previous page
105).
100).
(see "14.13.2 Fitting
98).
100).
OXYWAY
EN
105
Next page