Weinmann OXYWAY Fix I Service Manual page 138

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fast II
409
138
EN
OXYWAY
Back
5. Push the counter cap 404 over the safety valve and the housing.
6. Place the torque wrench onto the square of the counter cap 404.
228
7. Use the torque wrench to loosen the housing from the connecting
404
8. Remove the torque wrench and counter cap 404.
9. Unscrew the housing from the connecting bolt 228 by hand and
10.Use the watchmaker's screwdriver 409 to remove the rib filter 204
204
228
15.12.2 Fitting the rib filter
1. Insert the rib filter 204 into the connecting bolt 228.
204
228
2. Screw the housing onto the connecting bolt 228 by hand.
3. Push the counter cap 404 over the safety valve and the housing.
4. Place the torque wrench onto the square of the counter cap 404.
5. Use the torque wrench to tightly screw the housing onto the
404
228
6. Remove the torque wrench and counter cap 404.
7. Loosen the hand-screwed connection 230 at the top part of
8. Unscrew the hand-screwed connection 230 from the bottom
9. Remove the hand-screwed connection 230 of the pressure
10.Remove the top part of the counter tool 400 from the bottom
Forward
bolt 228.
then remove.
from the connecting bolt 228.
connecting bolt 228.
the counter tool 400.
part of the counter tool 400 by hand.
reducer from the counter tool 400.
part of the counter tool 400.