Weinmann OXYWAY Fix I Service Manual page 188

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fine I
267
188
EN
OXYWAY
Back
16.15.2 Fitting the hexagon nut
1. Tightly screw in the hexagon nut 268 using the socket wrench.
268
2. Fit the cover for the multistep flow selector 269
16.16 Replacing the multistep flow selector
16.16.1 Removing the multistep flow selector
Requirement
• The cover for the multistep flow selector 269 has been removed
• The hexagon nut 268 has been removed
1. Remove the multistep flow selector 267 from the housing cover 264
16.16.2 Fitting the multistep flow selector
1. Place the multistep flow selector 267 on the housing cover 264.
2. Fit the hexagon nut 268
3. Fit the cover for the multistep flow selector 269
16.17 Replacing the driving plate
16.17.1 Removing the driving plate
Requirement
• The cover for the multistep flow selector 269 has been removed
• The hexagon nut 268 has been removed
• The multistep flow selector 267 has been removed
Forward
When doing so, note: Hold the multistep flow selector 267 with
your hand during this step.
(see "16.14.2 Fitting the cover for the multistep flow selector",
page
187).
(see "16.14.1 Removing the cover for the multistep flow selector",
page
187).
(see "16.15.1 Removing the hexagon nut", page
by hand.
When doing so, note: The driving plate 266 must be inserted in
the groove of the multistep flow selector 267.
(see "16.15.2 Fitting the hexagon nut", page
(see "16.14.2 Fitting the cover for the multistep flow selector",
page
187).
(see "16.14.1 Removing the cover for the multistep flow selector",
page
187).
(see "16.15.1 Removing the hexagon nut", page
(see "16.16.1 Removing the multistep flow selector", page
Contents
Previous page
187).
188).
187).
188).
Next page