Final inspection OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
32
EN
OXYWAY
Back
8.4 Checking the system for leaks
1. Screw the hand-screwed connection 177 onto the valve of
177
2. Screw the hand-screwed connection 177 tight by hand.
159
3. Set the multistep flow selector 159 to "0".
4. Open the valve of the oxygen cylinder slowly.
5. Use a brush to lightly moisten the connection between the
6. Use a brush to lightly moisten the connection between the contents
7. Use a brush to lightly moisten the safety valve with leak detection
8. Use a brush to lightly moisten the small opening between the
9. Screw the cap 504 onto the inhalation outlet 155.
10.Set the multistep flow selector 159 to "15".
11.Use a brush to lightly moisten the connection between the
12.Set the multistep flow selector 159 to "0".
13.Remove the cap 504 from the inhalation outlet 155.
14.Close the valve on the oxygen cylinder.
Forward
the oxygen cylinder.
connecting bolt 175 and the pressure reducer housing with
leak detection spray 502.
Requirement:
No bubbles form between the connecting bolt 175 and the
housing.
indicator 150 and the housing with leak detection spray 502.
Requirement:
No bubbles form between the contents indicator 150 and the
housing.
spray 502.
Requirement:
No bubbles form on the safety valve.
housing and the pressure reducer housing cover 162 with leak
detection spray 502.
Requirement:
No bubbles form at the small opening between the housing and
the housing cover 162.
multistep flow selector 161 and the inhalation outlet 155 with
leak detection spray 502.
Requirement:
No bubbles form between the multistep flow selector 161 and
the inhalation outlet 155.