Weinmann OXYWAY Fix I Service Manual page 196

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fine I
264
264
196
EN
OXYWAY
Back
16.23 Replacing the spindle
16.23.1 Removing the spindle
Requirements
• The cover for the multistep flow selector 269 has been removed
• The hexagon nut 268 has been removed
• The multistep flow selector 267 has been removed
• The driving plate 266 has been removed
• The assembly housing cover 275 has been removed
• The spring for the piston 262 has been removed
• The piston 261 has been removed
• The spring for the lifting pin 263 has been removed
• The lifting pin with shoulder 265 has been removed
1. Remove the spindle 270 from the housing cover 264 by hand.
270
16.23.2 Fitting the spindle
1. Insert the spindle 270 into the housing cover 264.
270
2. Fit the lifting pin with shoulder 265
3. Fit the spring for the lifting pin 263
4. Fit the piston 261
Forward
(see "16.14.1 Removing the cover for the multistep flow selector",
page
187).
(see "16.15.1 Removing the hexagon nut", page
(see "16.16.1 Removing the multistep flow selector", page
(see "16.17.1 Removing the driving plate", page
(see "16.18.1 Removing the assembly housing cover", page
(see "16.19.1 Removing the spring for the piston", page
(see "16.20.1 Removing the piston", page
(see "16.21.1 Removing the spring for the lifting pin", page
(see "16.22.1 Removing the lifting pin with shoulder", page
(see "16.22.2 Fitting the lifting pin with shoulder", page
(see "16.21.2 Fitting the spring for the lifting pin", page
(see "16.20.2 Fitting the piston", page