Repairs to OXYWAY Click
208
EN
OXYWAY
Back
17.4.2 Attaching a label
1. Pull the protective film off the back of the label 321.
2. Stick the label 321 on the pressure reducer at the previously used
321
17.5 Replacing the pressure screw in the
inhalation outlet
Required tools
Slot-head screwdriver
17.5.1 Removing the pressure screw from the
inhalation outlet
1. Use the slot-head screwdriver to loosen the pressure screw 302 in
2. Remove the pressure screw 302 from the inhalation outlet 301.
301
302
17.5.2 Fitting the pressure screw in the inhalation
outlet
1. Insert the pressure screw 302 into the inhalation outlet 301.
301
302
2. Use the slot-head screwdriver to tightly screw in the pressure
Forward
When doing so, note: Do not touch the adhesive surface of the
label 321.
position.
When doing so, note: The adhesive surface must be free from dust
and grease.
the inhalation outlet 301.
When doing so, note: The slot on the pressure screw 302 must be
facing upwards when the screw is fitted.
screw 302 in the inhalation outlet 301.