Weinmann OXYWAY Fix I Service Manual page 150

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fast II
234
235
213
220
219
211
221
234
150
EN
OXYWAY
Back
4. Place the cover assembly 234 onto the orifice plate assembly 235.
5. Remove the cover assembly 234 from the clamping device for cover
6. Align the orifice plate assembly 235 in the cover assembly 234.
7. Carefully place the cover assembly 234 and the orifice plate
8. Insert the needle 221.
9. Insert the sliding ring 216.
10.Insert the springs 217.
11.Place the balls 215 on the springs 217.
12.Place the multistep flow selector 211 on the cover assembly 234
216
215
217
Forward
assembly 417 together with the orifice plate assembly 235.
When doing so, note: The bores in the multistep flow selector 213,
the bore in the orifice plate 220, and the grooved pin in the holder
for the orifice plate 219 must be in a straight line.
assembly 235 back onto the clamping device for cover
assembly 417.
When doing so, note: The position of the orifice plate assembly 235
must not change.
with the needle 221, sliding ring 216, springs 217 and balls 215.
When doing so, note: The "0" on the multistep flow selector 211
must be at the level of the needle 221.