Weinmann OXYWAY Fix I Service Manual page 29

Pressure reducer
Table of Contents

Advertisement

Accessory
Oxygen cylinder
Pressure hose
Connection nozzle
Instructions for Use
Back
7.9 Checking the accessories
1. Also check the accessories according to the following table:
Requirement
• The oxygen cylinder shows no signs of external damage.
• The valve of the oxygen cylinder is fully functional.
• The pressure reducer on the oxygen cylinder is fully functional.
• The oxygen cylinder has a valid test certificate.
• The pressure hose shows no signs of external damage.
• The O-ring in the pressure hose shows no signs of external damage.
• The pressure hose label is present and legible.
• The connection nozzle shows no signs of external damage.
• The O-ring in the connection nozzle shows no signs of external damage.
The instructions for use are present.
Forward
Final inspection OXYWAY Fix I, OXYWAY Fix III