Weinmann OXYWAY Fix I Service Manual page 53

Pressure reducer
Table of Contents

Advertisement

Accessory
Oxygen cylinder
Connection nozzle
Instructions for Use
Back
14.Read off the flow value at the testing device 510 and enter in
15.Close the multistep flow selector 267 until the flow indicator 251
16.Remove the special kit 500 from the testing device 510.
17.Connect the special kit 500 to the testing device 512.
18.Read off the flow values at the testing device 512 and enter in
19.Close the valve on the oxygen cylinder.
20.Open the multistep flow selector 267 until the flow indicator 251
21.Close the multistep flow selector 267.
22.Remove the special kit 500 from the testing device 512.
23.Remove the special kit 500 from the adapter 507
24.Remove the adapter 507 from the inhalation outlet 257.
25.Remove the hand-screwed connection 273 from the valve of
10.9 Checking the accessories
1. Also check the accessories according to the following table:
Requirement
• The oxygen cylinder shows no signs of external damage.
• The valve of the oxygen cylinder is fully functional.
• The pressure reducer on the oxygen cylinder is fully functional.
• The oxygen cylinder has a valid test certificate.
• The connection nozzle shows no signs of external damage.
• The O-ring in the connection nozzle shows no signs of external damage.
The instructions for use are present.
Forward
the test report.
Requirement:
The flow value is 3 l/min
±
displays 0 l/min.
the test report.
Requirement:
The flow value is 0 l/min ± 0.04 l.
displays 15 l/min.
the oxygen cylinder.
Contents
Previous page
Final inspection OXYWAY Fine I
0.5 l/min.
OXYWAY
Next page
EN
53