15.6 Replacing the sieve and nozzle plate
in the inhalation outlet
Required tools
• Watchmaker's screwdriver 409
• Tweezers
15.6.1 Removing the sieve and nozzle plate from the
inhalation outlet
Requirement
The pressure screw 206 has been removed from the inhalation
outlet 205
inhalation outlet", page
1. Use the watchmaker's screwdriver 409 to carefully remove the
sieve 207 and sealing disks 208 from the inhalation outlet 205.
15.6.2 Fitting the sieve and sealing disks in the
inhalation outlet
1. Use tweezers to insert the sealing disk 208 in the inhalation
outlet 205.
2. Use tweezers to insert the sieve 207 in the inhalation outlet 205.
3. Use tweezers to insert the sealing disk 208 in the inhalation
outlet 205.
4. Fit the pressure screw 206 in the inhalation outlet 205
(see "15.5.2 Fitting the pressure screw in the inhalation outlet",
page
15.7 Replacing the pressure screw in the
pressure outlet
Required tools
Slot-head screwdriver
15.7.1 Removing the pressure screw from the pressure
outlet
1. Use the slot-head screwdriver to loosen the pressure screw 210
in the pressure outlet 209.
2. Remove the pressure screw 210 from the pressure outlet 209.
209
210
Contents
(see "15.5.1 Removing the pressure screw from the
132).
132).
Previous page
Repairs to OXYWAY Fast II
OXYWAY
EN
133
Next page