Weinmann OXYWAY Fix I Service Manual page 192

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Fine I
262
192
EN
OXYWAY
Back
16.19.2 Fitting the spring for the piston
1. Position the spring 262 on the piston 261.
261
2. Insert the piston 261 into the assembly housing cover 275.
3. Fit the assembly housing cover 275
4. Fit the driving plate 266
5. Fit the multistep flow selector 267
6. Fit the hexagon nut 268
7. Fit the cover for the multistep flow selector 269
16.20 Replacing the piston
Required tools
Cloth
16.20.1 Removing the piston
Requirement
• The cover for the multistep flow selector 269 has been removed
• The hexagon nut 268 has been removed
• The multistep flow selector 267 has been removed
• The driving plate 266 has been removed
• The assembly housing cover 275 has been removed
• The spring for the piston has been removed
1. Remove the piston 261 from the assembly housing cover 275.
2. Remove the spring for the lifting pin 263 from the piston 261.
Forward
(see "16.18.2 Fitting the assembly housing cover", page
(see "16.17.2 Fitting the driving plate", page
(see "16.16.2 Fitting the multistep flow selector", page
(see "16.15.2 Fitting the hexagon nut", page
(see "16.14.2 Fitting the cover for the multistep flow selector",
page
187).
(see "16.14.1 Removing the cover for the multistep flow selector",
page
187).
(see "16.15.1 Removing the hexagon nut", page
(see "16.16.1 Removing the multistep flow selector", page
(see "16.17.1 Removing the driving plate", page
(see "16.18.1 Removing the assembly housing cover", page
(see "16.19.1 Removing the spring for the piston", page 191)
Contents
Previous page
190).
189).
188).
188).
187).
188).
188).
189).
Next page