15.30 Disassembling/assembling the cover
assembly
Required tools
• Cover clamping device 401
• Vise
• Special counter tool, DISS 416
• Special gripping pliers 415
• Counter tool with fastening handle 400
• Counter cap 404
• Open-end wrench A/F 15
• Watchmaker's screwdriver 409
• Brush
• Oxygen-compatible lubricant 414
• Allen key size 5
• Thread tap G 1/8
• Brass brush
• Torque wrench 8 Nm ± 2 Nm
• Torque wrench 16 Nm ± 1 Nm
• Socket wrench size 15
• Loctite 413
15.30.1 Disassembling the cover assembly
Requirement
• The multistep flow selector assembly 233 has been removed
(see "15.16.1 Removing the multistep flow selector assembly",
page
• The spring for the piston 227 has been removed
(see "15.17.1 Removing the spring for the piston", page
• The piston 226 has been removed
(see "15.19.1 Removing the piston", page
• The multistep flow selector assembly 233 has been disassembled
(see "15.20.1 Disassembling the multistep flow selector assembly",
page
1. Lightly tension the cover clamping device 401 in the vise.
2. Place the cover assembly 234 in the cover clamping device 401.
3. Align the cover assembly 234 in the cover clamping device 401
234
so that the cover clamping device 401 clamps onto the housing
214
cover 214.