Weinmann OXYWAY Fix I Service Manual page 151

Pressure reducer
Table of Contents

Advertisement

417
211
Back
Forward
13.Fix the multistep flow selector 211 in place using the clamping
device for cover assembly 417.
14.Use the Allen key to tightly screw in the threaded pin 212
on the multistep flow selector 211.
15.Loosen the fastening on the clamping device for cover
assembly 417.
16.Remove the multistep flow selector assembly 233 from the
clamping device for cover assembly 417.
17.Check the alignment of the orifice plate assembly 235 in the
multistep flow selector assembly 233.
18.If necessary: Re-align the orifice plate assembly 235. To do so,
disassemble the multistep flow selector assembly again and then
re-assemble.
19.Fit the piston 226
20.Fit the spring for the piston 227
(see "15.17.2 Fitting the spring for the piston", page
21.Fit the multistep flow selector assembly 233
(see "15.16.2 Fitting the multistep flow selector assembly",
page
15.21 Replacing the O-ring in the contents
indicator
Required tools
Watchmaker's screwdriver 409
15.21.1 Removing the O-ring from the contents
indicator
Requirement
• The multistep flow selector assembly 233 has been removed
(see "15.16.1 Removing the multistep flow selector assembly",
page
• The threaded locking pin 201 has been removed
(see "15.18.1 Removing the threaded locking pin", page
• The contents indicator 200 has been removed
(see "15.22.1 Removing the contents indicator", page
1. Use the watchmaker's screwdriver 409 to carefully remove
the O-ring 202 from the housing.
Contents
(see "15.19.2 Fitting the piston", page
145).
144).
Previous page
Repairs to OXYWAY Fast II
148).
146).
147).
152).
OXYWAY
EN
151
Next page