17.3 Assemblies
This chapter contains references to the assemblies defined in the
following table.
Assembly
Multistep flow selector
assembly 323
Cover assembly 324
Orifice plate assembly 325
The assemblies are not available as complete units; only as individual
parts of the assemblies.
To remove parts other than those listed in the table, the assemblies can
be removed as complete units.
17.4 Replacing a label
Required tools
• Knife
• Spirit or Label remover
17.4.1 Removing a label
1. Remove the label 321 from the pressure reducer.
2. Use spirit or label remover to remove any residue from the pressure
321
reducer.
3. Dispose of the label 321.
Contents
Repairs to OXYWAY Click
Parts
• Multistep flow selector
• Threaded pin
• 2 balls
• 2 springs
• Sliding ring
• Needle
• Orifice plate assembly
• Cover assembly
• Housing cover
• O-ring in the housing cover
• Union nut
• Multistep flow selector
• 2 O-rings in the multistep flow selector
• Inhalation outlet
• Holder for the orifice plate
• Orifice plate
• Slide disk